The HiSeq and dedicated workstation have the following dimensions after installation:
Measurement |
Instrument |
Workstation |
Width |
118.6 cm (46.7 in.) The waste bottle extends an additional |
21.6 cm (8.5 in.) |
Height |
94 cm (37 in.) Minimum height The instrument height is adjustable to an additional 1.27 cm (0.5 in.). |
57 cm (22.3 in.) |
Depth |
76 cm (30 in.) The keyboard tray extends an additional |
55.4 cm (21.8 in.) |
Weight |
221.4 kg (488 lbs.) |
34 kg (75 lbs.) |
Type |
Specification |
Line Voltage |
100 -240 Volts AC @ 50/60 Hz |
Power Consumption |
Maximum 1500 Watts |
Bench-Craft (www.bench-craft.com), part #HS-30-60-30 P2with casters.
Width |
Height |
Depth |
Locking Casters |
152.4 cm (60 in.) |
76.2–91.4 cm (30–36 in.) |
76.2 cm (30 in.) |
Yes |
Item |
Source |
Balance, top-loading, digital, 420 g capacity |
缝匠肌、模型CPA423S (or similar) |
Bottle rack, 6-position 250 ml centrifuge tube, epoxy-coated |
LabScientific, catalog # CBR 200 |
Freezer, -15° to -25°C |
General lab supplier |
Ice bucket |
General lab supplier |
Refrigerator, 2° to 8°C |
General lab supplier |
Tweezers, square-tip plastic |
McMaster-Carr, catalog # 7003A22 |
Consumable |
Supplier |
Purpose |
Tween 20, viscous liquid |
Sigma-Aldrich, catalog # P7949 |
Instrument maintenance wash. |
ProClin 300, 50 ml | Sigma-Aldrich, catalog # 48912-U | Instrument maintenance wash. |
Alcohol wipes, or Ethanol, 70% |
VWR, catalog # 95041-714 General lab supplier |
Cleaning the flow cell and flow cell stage. |
Centrifuge tube, 250 ml |
Corning, catalog # 430776 |
Instrument and monthly maintenance wash. |
Conical tubes, 15 ml |
Corning, catalog # 430052 |
Collecting and measuring waste volumes. |
Conical tubes, 50 ml, |
Corning, catalog # 430921 |
Storing flow cells. |
Carboy, at least 6 liters | General lab supplier | Maintenance wash solution. |
Disposable gloves, powder-free |
General lab supplier |
General use. |
Lab tissue, low-lint |
VWR, catalog # 21905-026 |
Cleaning the flow cell holder. |
Lens paper, 4 x 6 in. |
VWR, catalog # 52846-001 |
Cleaning the flow cell. |
Pipette tips, 200 µl |
General lab supplier |
Preparing and aliquoting reagents. |
Pipette tips, 1000 µl |
General lab supplier |
Preparing and aliquoting reagents. |
Water, laboratory-grade |
Millipore |
SBS reagent rack, position 2. Instrument wash. |
For more information, see theHiSeq Site Preparation Guide.